SimXray-CR Anatomically Programmed Interface (API)

 
  • milliampere values may be selected in the range of 100 - 1200 in increments as small as 25 milliamperes.
  • mSec values may be selected in a range of 1 - 10000 in increments as small as 1 millisecond.
  • mSec values are not available when AEC is active.
  • kVp values may be selected in a range of 25 - 150 in increments as small as 1 kVp.
  • Technical Factors are automatically set when a Projection is selected.
  • Technical Factors are selected from the built-in Technique Chart.